Kuhner Lab-Therm LT-XC Incubator Shaker, Dual Stack with Cooling Option
- Make Kuhner
- Hertz 50Hz/60Hz
- Model Lab-Therm LT-XC
- Serial 67001.972 & 67001.871
- Weight 788
- Voltage 120V
- Working Working
- Includes Power Cord
- Warranty 30-Day Warranty, 100% Parts and Labor
- Dimensions 347×38×73 in
- Shipping Type Freight
- Item Condition Pre-owned
Kuhner Lab-Therm LT-XC Incubator Shaker, Dual Stack with Cooling Option
***Dual stack incubators was tested to specifications. Heating, cooling, UV, shakers all working. (Note: Top incubator does not have refrigeration module installed.) Unit is preowned with minimal wear from normal use, (scratches on left side, chipped paint on bottom incubator.) 30-day warranty included with purchase. See photos for details.
Space saving stackable incubator shaker
The Kuhner LT-X (Lab-Therm) incubator shaker has the same capabilities as the larger ISF1-Z and ISF4-X models. The incubator shaker was specially developed to address the shaking needs of the biotechnology and pharmaceutical industries. The practical size of the LT-X means it will fit in any laboratory.
-
Fits into any laboratory
-
Used in biotech and pharmaceutical industries
-
Accepts flasks up to 6 litres
-
Two units can be stacked without the need for special kits or tools
-
Humidity, CO2 and O2 control available
Extra space
Exceptional chamber height means flasks up to 6 litres capacity can be accommodated. If using smaller flasks, there is sufficient head height to fit one or two stationary trays above the shaking table. It is also possible to stack two LT-X together without the need for special kits or tools.
Functionality through direct drive
The maintenance-free direct drive guarantees long and economical performance from the LT-X. The shaking diameter can be changed at any time thanks to Kuhner’s unique direct drive system. The integrated drive is protected from spills and the all stainless steel interior has a drain at the back for any unwanted liquids. The chamber can be cleaned using chemicals or sterilized with optional UV lighting.
Safety and quality
A door switch ensures the motor stops as soon as the door is opened. After closing the door properly, the motor restarts.
Configuration
The LT-X can be equipped with options for CO2, O2,controlled humidity, LED photosynthesis, UV lighting and much more. It is now also possible to integrate measuring data from Kuhner shakers with your company’s own network using Kuhner Insight software together with the NET-60 interface.
Kuhner shakers are suitable for the cultivation of microbial, plant, human and animal cell cultures. Find more information about the different application areas and suitable product recommendations for optimal cultivation here.
Technical Overview:
LT-X(C) Basic |
LT-X(C) EcoDew® |
LT-X(C) Compressor |
|
Cooling |
no |
no |
yes |
Humidity control |
no |
yes |
no / yes |
Temperatur minimal |
ambient + 10 °C |
ambient + 10 °C |
ambient – 15 °C (– 10 °C)* |
Temperatur maximal |
80°C (60°C)* |
80°C (60°C)* |
80°C (60°C)* |
Humidity maximal |
- |
85 % r.h. |
- / 85 % r.h. |
Power consumption |
< 700W |
< 1050W |
< 880W / < 1150W |
Machine:
Gas volume |
260 l |
Weight (with cooling) |
170 kg |
Illumination |
LED |
Operating menu in |
German, French, English, Italian, Spanish |
Interface, standard |
CAN-Bus |
Interface, optional |
USB, Ethernet, digital, analogue |
Ambient temperature |
10°C up to 35°C |
Temperature :
Setting, digital |
0.1°C |
Accuracy, absolute |
± 0.30°C (37°C) ± 0.25°C (37°C)* |
Principle of sensor |
Pt-100 |
Power of heating |
500 W |
Power of cooling |
90...155 W |
Air circulation |
160 m3/h |
max. at 25..55°C |
85% r.h. |
Setting, digital |
1% r.h. |
Accuracy, absolute |
± 2% r.h. |
Principle of sensor |
capacitive |
Water refill |
automatic |
Water heater |
180 W |
Door heater |
90 W |
CO2:
Principle of sensor |
Infrared, NDIR |
Measuring range |
0...20% CO2 |
Setting, digital |
0.1% |
Accuracy, absolute (including non-linearity, calibration uncertainty, repeatability) |
± 0.40% at 5% CO2 |
Temperature range |
5...60°C |
CO2-supply |
max. 2 bar overpressure |
O2:
O2 sensor |
zirconium oxide |
Measuring range |
approximately 0% up to ambient O2 (20.95% at dry air) |
Accuracy, absolute |
±0.40% at 5% O2 |
O2 sensor stability
|
±0.5% full scale/year |
Setting, digital |
± 0.1% |
O2 control |
PID, digital |
Temperature range |
-10...80°C |
N2-supply |
max. 0.5...0.8 bar overpressure |
Mains Connection:
110 - 120 V / 50-60 Hz |
|
95 - 105 V / 50-60 Hz |
Shaking Unit:
Tray size |
EX (500 x 420mm) |
Loading, maximum |
25kg |
Setting, digital |
1 rpm |
Accuracy, absolute |
≤ 800 rpm = ± 0.5 rpm |
Timer |
1s ... 999h |
Acceleration |
controlled |
Active brake |
adjustable |
Stop on position |
adjustable |
Shaking Motion:
|
Speed |
* orbital, Ø 12.5mm |
20...500 rpm |
* orbital, Ø 25.0mm |
20...400 rpm |
* orbital, Ø 50.0mm |
20...300 rpm |
* linear, ↔ 12.5mm |
20...400 rpm |
* linear, ↔ 25.0mm |
20...300 rpm |
* linear, ↔ 50.0 mm |
20…200 rpm |