GE Healthcare ImageQuant LAS 4000 Luminscent Image CLEARANCE! As-Is
- Make GE Healthcare
- Hertz 50Hz/60Hz
- Model ImageQuant LAS 4000
- Serial 2689092
- Weight 220
- Voltage 120V/220V
- Working Working
- Includes Power Cord
- Warranty AS-IS
- Dimensions 23×24×42 in
- Shipping Type Freight
- Item Condition Pre-owned
- Manufacturing Date 2012
GE Healthcare ImageQuant LAS 4000 Luminscent Image Analyzer
**This instrument has only been tested for basic power function. This item has slight visible scratches on bottom front panel and side of unit from normal use. Does NOT include ImageQuant Software. See pictures for further details. This unit will need software and calibration by buyer.
The GE ImageQuant LAS-4000 Multi-Mode Imager is a robust digital imaging system of gel or membrane samples or films. This imaging system uses epi or tans light on gels stained by EtBr and exposes a sample repeatedly, incrementally, or by programming exposure schemes. This flexible system is modular and can use a combination of light sources and filters to achieve the technique you need, including upgrading to an IR fluorescence capability, available from GE.
The wide linear dynamic range of the LAS-4000 avoids multiple film exposures by accurately quantitating both weak and strong signals over 4 orders of magnitude. The F0.85 43 mm LAS high sensitivity lens of this quantitative biomolecular imager has a 21 × 14 cm field of view with a 6.3-megapixel image resolution thanks to the innovative diagonal pixel configuration. This camera is cooled to -25°C (maximum -35°C) allowing users longer exposure times so theres less background and yielding more precise quantitation of weak signals.
The GE ImageQuant LAS-4000 Multi-Mode Imager is capable of Quantitative Western blotting, UV fluorescence, and Infrared fluorescence when used with the right reagents, optional colored epi illuminators, and filters installed. This system is comprised of a dark box, camera, lens, and a five-position filter wheel that gives users the flexibility to change methods. To control this system the robust ImageQuant LAS 4000 software easily guides users through programming Focusing, filters, illuminators, and exposure time without needing to open the cabinet. The automated iris of this imaging system responds quickly to changes in focus and other settings for quick turnaround.
Manufacturers Specifications:
Emission Filter | 605DF40 EtBr EtBr detection |
Lens | FUJINON F0.85 43 mm LAS high sensitivity lens 21 × 14 cm |
Cooling | Two-stage thermoelectric module with air circulation |
Cooling temperature | Down to -35°C (at room temp. -28°C) |
Dynamic range | 16-bit, 4 orders of magnitude |
Image resolution | Maximum 3072 × 2048, 6.3 Mpixels |
Pixel size Approximately | 11 ?m |
Focusing and aperture | Automatic, remote operation |
Capture modes | Automatic, manual (normal/incremental/repetitive/program) |
Exposure time | Automatic, manual (1/100 s to 30 h) |
Pixel correction | Dark frame correction, flat frame correction, and distortion correction |
Image quality correction | Binning, smoothing |
Binning modes | 1 × 2, 2 × 4, 4 × 8, 8 × 16 |
Image size | Up to 12 MB (.GEL and .TIF) |
Sample size | 21 × 14 cm (25 × 25 cm with SIGMA wide view lens) |
Environmental Tolerances | 15°C to 28°C, Humidity 30% to 70% (no condensation) |
Interface | USB 2.0 |
Testimonials
“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”
“Great company to work with. Tammy completed our first international equipment shipment with ease.”
“Fast response, open to adjusting schedule as needed, and great customer interaction.”