Kaye Instruments LTR-25/140 Dry Block Calibrator LTR-140
- Make Kaye Instruments
- Model LTR-140
- Serial A4B944H
- Weight 38 lbs
- Working Working
- Includes Power Cord
- Warranty 30-Day Warranty, 100% Parts and Labor
- Dimensions 20 × 17 × 21 in
- Shippingtype UPS Ground
- Item Condition Used
- Post Views Count 1234
The LTR-25/140 temperature reference dry well is specifically designed to fully automate sensor calibration when used with Kaye validation software and traceable temperature standards.
Kaye dry wells employ unique inserts which minimize cooling of the thermocouple tips due to heat conduction providing an uncertainty of merely 0.1°C. Without proper inserts, transfer uncertainty in excess of 0.5°C can occur with 22 gage, type T thermocouples.
Manufacture Specifications
Temperature Range | -25°C to 140°C |
Ambient Operating Range | 5°C to 50°C |
Set-Pointing Accuracy | 0.2°C |
Temperature Stability | 0.02°C |
Transfer Calibration Accuracy* IRTD Standard to Thermocouples |
–25°C to 80°C: ±0.1°C 80°C to 130°C: ±0.15°C 130°C to 140°C: ±0.18°C |
Typical Heat Up Time | 25°C to 80°C: 6 minutes 25°C to 140°C: 14 minutes |
Typical Cool Down Time | 10 minutes: 25°C to –20°C |
Display | LED w/0.01°C resolution |
Computer Interface | RS232 |
Testimonials
“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”
“Great company to work with. Tammy completed our first international equipment shipment with ease.”
“Fast response, open to adjusting schedule as needed, and great customer interaction.”