The Agilent 1200 HPLC G1321A Fluorescence Detector brings high-sensitivity and selection through simultaneous multi-wavelength detection. This easy-to-use detector provides quantitative data and fluorescence spectra from a single run. The FD's spectral information can be used for rapid method optimization and verification of separation quality.
Manufacture Specification
Detection Type |
Multi-signal fluorescence detector with rapid on-line scanning capabilities and spectral data analysis |
Performance Specifications |
10 fg Anthracene, Ex=250 nm, Em=400 nm* RAMAN single wavelength (H2O) > 500 with Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell time constant=4 seconds (8 seconds responsetime) RAMAN dual wavelength (H2O) > 300 with Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell time constant=4 seconds (8 seconds responsetime) |
Light source |
Xenon Flash Lamp, normal mode 20 W, economy mode 5 W |
Pulse frequency |
296 Hz for single signal mode 74 Hz for spectral mode |
Excitation Monochromator |
Range: 200 nm - 700 nm and zero-order Bandwidth: 20 nm (fixed) Monochromator:concave holographic grating, F/1.6, blaze: 300 nm |
Emission Monochromator |
Range: 280 nm - 900 nm and zero-order Bandwidth: 20 nm (fixed) Monochromator:concave holographic grating, F/1.6, blaze: 400 nm |
Reference System: |
in-line excitation measurement |
Timetable programing |
up to 4 signal wavelengths, response time, PMT Gain, baseline behavior (append, free, zero), spectral parameters |
Spectrum acquisition: |
Excitation or Emission spectra Scan speed: 28 ms per datapoint (e.g. 0.6 s/spectrum 200-400 nm, 10 nm step) Step size: 1-20 nm Spectra storage: All |
Wavelength characteristic |
Repeatability+/- 0.2 nm Accuracy+/- 3 nm setting |