Waters Acquity UPLC System
- Make Waters
- Hertz 50Hz/60Hz
- Model Acquity UPLC
- Serial 186015029, 186015006, 186015001
- Weight 219
- Voltage 120V/220V
- Working Working
- Includes Power Cord
- Warranty AS-IS
- Dimensions 36×24×46 in
- Shipping Type Freight
- Item Condition Pre-owned
The Waters Acquity UPLC system provides you with reproducible, dependable and data rich results for challenging separations. This UPLC system features a wide range of flow rates, back pressures and linear velocities, with increased throughput with high efficiency so you get the most for your time and resources. This contained tower uplc system holds a sample manager, column heater, binary solvent manager, and a PDA detector. This ultra-performance liquid chromatography system has a unique sub 2 μm particle system that offers 3 times the efficiency of a 5 μm system with 70% higher resolution. This higher resolution means more information from more samples in less time. The Waters UPLC sytem can then run separations at a flow rate 3 times faster than HPLC, without a loss in resolution. This Aquity system contains a PDA Detector, sampler manager autosampler, a binary solvent manager, as well as a solvent and waste tray.
Manufacture Specification
Solvent Manager |
|
Number of solvents |
Up to four, in any combination of two: A1 or A2 and B1 or B2. |
Solvent conditioning |
Vacuum degassing (six-channel): one channel per solvent, and two channels for sample manager wash solvents. |
Settable flow rate range |
0.001 to 2.000 mL/min, in 0.001 mL increments |
Gradient profiles |
Eleven gradient curves [including linear, step (2), concave (4), and convex(4)] Wet prime Automatic
|
Maximum operating pressure |
103,421 kPa (1034 bar, 15,000 psi) up to 1 L/min, 62,053 kPa (621 bar, 9000 psi) up to 2 mL/min |
Composition accuracy |
±0.5% absolute (full scale) from 5 to 95% from 0.50 to 2.00 mL/mina |
Composition precision |
0.15% RSD or ±0.04 min SD |
Flow accuracy |
±1.0% at 0.5 mL/min |
Sample Manager |
|
Number of SBS/ANSI sample plates |
Total of two plates, expandable to up to 22 plates with optional sample organizer. Maximum of 22 microtiter plates, 11 intermediate depth plates, or 8 deep-well (or 2 mL vial) plates: 96 and 384 microtiter plates 48-position 2 mL vial plates 48-position 0.65 mL micro-centrifuge tube plates 24-position 1.5 mL micro-centrifuge tube plates |
Maximum sample capacity |
768 in two 384-well plates; expandable to up to 8,448 samples with optional sample organizer |
Injection volume range |
0.1 to 50 µL, in 0.1 µL increments, partial or full loop mode, 10 µL loop is standard; 1, 2, 5, 20, and 50 µL loops
|
Sample temperature control |
4 to 40 °C, programmable in 0.1 °C increments |
Minimum sample required |
5 µL residual volume at a needle offset of 0, or 22 µL at the preset 2 mm needle offset |
Column Heater |
|
Columns accepted |
Any up to 4.6 mm ID and 150mm long |
Temperature range |
5 above ambient to 65°C (9 above ambient to 194°F) |
PDA Detector |
|
Wavelength Range |
190-500 nm |
Flow Cell Options |
Analytical Cell 500 nanoliters, 10 mm pathlength High Sensitivity Cell 2.4 microliters, 25mm pathlength |
Testimonials
“REUZEit has been a great partner for our used equipment needs and always provide timely updates of new arrivals of consigned equipment.”
“Great company to work with. Tammy completed our first international equipment shipment with ease.”
“Fast response, open to adjusting schedule as needed, and great customer interaction.”